<div class="csl-bib-body">
<div class="csl-entry">Wobrauschek, P., Holub, E., & Kregsamer, P. (2022, June 28). <i>Spectral Correction for Transmission X-ray Tubes for Use in the Quantification Software ATI-QUANT</i> [Conference Presentation]. EXRS 2022 - European X-ray Spectrometry Conference, Brügge, Belgium. http://hdl.handle.net/20.500.12708/136307</div>
</div>
-
dc.identifier.uri
http://hdl.handle.net/20.500.12708/136307
-
dc.description.abstract
ATI-QUANT is a software package developed at TU Wien, Atominstitut[1,2]. ATI-QUANT uses fundamental parameters for quantification in XRF. Considerable efforts were made to describe the spectral distribution of the transmission X-ray tube AMPTEK MINI-X with Rh anode versus the previously used side-window tube with thick Rh target. The following research objectives were addressed: to analyze possible reasons for the incorrect calculation of elemental concentrations measured with the rhodium-anode transmission tube; to find a corrective function to better reflect the spectral distribution of the transmission tube; to test the corrected spectrum by means of reference standard materials; to extend the experimental setup in such a way as to correctly quantify light elements; and to compare two software packages for data evaluation (AXIL-IAEA and PyMCA).
By modifying the experimental setup and by calculation of the spectral distribution, the range of quantifiable elements was successfully extended to include light elements (down to Al). This was achieved by flushing the excitation and detection paths with He gas under slight overpressure. The corrected transmission-tube spectrum provides excellent quantitative results, achieving an even better performance than the well-established side-window spectrometer. Results and applications will be shown.
[1] B.Grossmayer, Softwarepaket zur quantitativen Röntgenfluoreszenzanalyse mittels fundamentaler Parameter unter Verwendung von Röntgenoptiken, Diploma Thesis, TU Wien, 2009
[2] P. Necker, Quantitative X-ray fluorescence analysis of samples with various matrices using a universal data evaluation software, Diploma Thesis, TU Wien, 2017
en
dc.language.iso
en
-
dc.subject
Quantification in XRF
en
dc.title
Spectral Correction for Transmission X-ray Tubes for Use in the Quantification Software ATI-QUANT
en
dc.type
Presentation
en
dc.type
Vortrag
de
dc.type.category
Conference Presentation
-
tuw.researchTopic.id
M2
-
tuw.researchTopic.name
Materials Characterization
-
tuw.researchTopic.value
100
-
tuw.publication.orgunit
E141-05 - Forschungsbereich Radiation Physics
-
tuw.author.orcid
0000-0002-3699-3003
-
tuw.event.name
EXRS 2022 - European X-ray Spectrometry Conference
en
tuw.event.startdate
26-06-2022
-
tuw.event.enddate
01-07-2022
-
tuw.event.online
On Site
-
tuw.event.type
Event for scientific audience
-
tuw.event.place
Brügge
-
tuw.event.country
BE
-
tuw.event.institution
University of Antwerp
-
tuw.event.presenter
Kregsamer, Peter
-
wb.sciencebranch
Chemie
-
wb.sciencebranch
Physik, Astronomie
-
wb.sciencebranch.oefos
1040
-
wb.sciencebranch.oefos
1030
-
wb.sciencebranch.value
50
-
wb.sciencebranch.value
50
-
item.openairecristype
http://purl.org/coar/resource_type/c_18cf
-
item.openairecristype
http://purl.org/coar/resource_type/c_18cf
-
item.grantfulltext
none
-
item.fulltext
no Fulltext
-
item.openairetype
Presentation
-
item.openairetype
Vortrag
-
item.cerifentitytype
Publications
-
item.cerifentitytype
Publications
-
item.languageiso639-1
en
-
crisitem.author.dept
E141 - Atominstitut
-
crisitem.author.dept
E164-02-1 - Forschungsgruppe Prozessanalytik
-
crisitem.author.dept
E141-05 - Forschungsbereich Radiation Physics
-
crisitem.author.orcid
0000-0002-3699-3003
-
crisitem.author.parentorg
E130 - Fakultät für Physik
-
crisitem.author.parentorg
E164-02 - Forschungsbereich Umwelt-, Prozessanalytik und Sensoren