<div class="csl-bib-body">
<div class="csl-entry">Elenkov, M. (2018). <i>Detection of aortic hemodynamics in Electrical Impedance Tomography using functional regions of interest : investigation of in-vivo, in-silico and phantom measurements</i> [Diploma Thesis, Technische Universität Wien]. reposiTUm. https://doi.org/10.34726/hss.2018.53923</div>
</div>
-
dc.identifier.uri
https://doi.org/10.34726/hss.2018.53923
-
dc.identifier.uri
http://hdl.handle.net/20.500.12708/6150
-
dc.description
Abweichender Titel nach Übersetzung der Verfasserin/des Verfassers
-
dc.description.abstract
Herz-Kreislaufversagen tritt bei mehr als 20% der Intensivpatienten auf, was das Todesrisiko um das 12-fache erhöht. Zentrale hämodynamische Parameter wie Schlaganfallvolumen und Blutdruck (BP) haben sich als die besten Prädiktoren für verschiedene Herz-Kreislauf-Erkrankungen erwiesen, sind aber herausfordernd nicht invasiv zu überwachen. Die derzeitige Praxis für ihre Messung erfordert eine Katheterisierung, die mit gesundheitlichen Komplikationen und einer erhöhten Sterblichkeitsrate bei Patienten verbunden ist. Andere Methoden, obwohl nicht invasiv, versuchen, die zentrale Hämodynamik durch periphere Messungen zu approximieren und sind mit zusätzlichen Mängeln verbunden. Oszillometrische Methoden messen BP zu diskreten Zeitpunkten, während die Tonometrie zum Beispiel aufdringlich und sehr empfindlich auf Bewegungsartefakte reagiert. Die elektrische Impedanztomographie (EIT) ist eine bildgebende Methode, die das Potenzial hat, diese Herausforderungen zu bewältigen, indem sie die Gesundheit des Herz-Kreislauf-Systems kontinuierlich und nicht-invasiv bewertet. Die Rekonstruktion von EIT-Bildern ist jedoch ein komplizierter Prozess, der eine Vielzahl von Eingabeparametern beinhaltet, wobei verschiedene Kombinationen dieser Parameter unterschiedliche Bilder und damit unterschiedliche physiologische Parameter erzeugen. Während es Bemühungen gibt, diese Probleme anzugehen, sind Rekonstruktionsparameter und Analysemethoden für das hämodynamische Monitoring mit EIT noch nicht standardisiert. In dieser Arbeit wurde eine umfassende Analyse des EIT-Bildgebungsverfahrens für hämodynamische Messungen in einer Reihe von Experimenten von In-silico- bis In-vivo-Messungen durchgeführt. Die besten Rekonstruktionsparameter und Einstellungen für das hämodynamische EIT wurden in Simulationen ausgewählt und für die Bildrekonstruktion verwendet. Bei der Verwendung von EIT ist eine genaue Erkennung der Aorta für die Berechnung der hämodynamischen Parameter wichtig. Ein neuartiger Algorithmus zur Erkennung von Aortenpixeln in EIT-Bildern wird vorgestellt. Die Genauigkeit des Algorithmus wurde an Tierdaten mit intermodaler (Computertomographie, F1-Score 0,25) und intramodaler (Injektion von Kontrastmittel, F-1-Score 0,8125) Bodenwahrheit validiert. In einem nächsten Schritt wurde das EIT-Signal an 5 gesunden Probanden gemessen und der Ort der Aorta mit dem Algorithmus erfolgreich bestimmt. Durch die Ausnutzung des EKG-gesteuerten Ensembles wurde eine repräsentative Aorten-BP-Wellenform erfasst, die eine Detektion der BP-Impulsankunftszeit (PAT) ermöglichte. Es wurde eine starke negative lineare Korrelation zwischen dem mittleren arteriellen Blutdruck und dem mit EIT nachgewiesenen PAT beobachtet (r=-0,8593). Die Ergebnisse dieser Arbeit sind vielversprechend und deuten darauf hin, dass die EIT-Bewertung hämodynamischer Parameter eine brauchbare Alternative zu den derzeitigen Methoden sein könnte. Die EIT-Technologie befindet sich jedoch noch in der Entwicklungsphase und die verschiedenen verfügbaren Rekonstruktionsalgorithmen mit ihren Rekonstruktionseinstellungen müssen in größerem Umfang weiter erprobt werden, gefolgt von methodischen In-vivo-Studien, um den Weg in die klinische Anwendung zu ebnen. Eine auf den Ergebnissen dieser Arbeit basierende Tagungsarbeit wurde zur Präsentation auf der 19 Elektrische Impedanz-Tomographie-Konferenz angenommen.
de
dc.description.abstract
Cardiovascular failure occurs in more than 20% of intensive care unit patients, which increases the risk of death by 12 fold by itself. Central hemodynamic parameters like stroke volume and blood pressure (BP), have been shown to be the best predictors for various cardiovascular diseases, but are challenging to monitor non-invasively. The current practice for their measurement requires catheterization which is associated with health complications and elevated mortality rate in patients. Other methods, although non-invasive try to approximate central hemodynamics trough peripheral measurements and are associated with additional shortcomings. Oscillometric methods measure BP at discrete time instances, while tonometry, for example, is an obtrusive and very sensitive to movement artefacts. Electrical Impedance Tomography (EIT) is an imaging modality that has the potential to overcome these challenges, by assessing the health of the cardiovascular system continuously and non-invasively. The reconstruction of EIT images however, is a complicated process that involves a lot of input parameters, where different combinations of these parameters produce different images and consequently different physiological parameters. While there are efforts to address these issues reconstruction parameters and analysis methods for the purpose of hemodynamics monitoring with EIT are still not standardized. In this work a comprehensive analysis of the EIT imaging process for hemodynamic measurements was made in a series of experiments from in-silico to in-vivo measurements. The best performing reconstruction parameters and settings for hemodynamics EIT were selected in simulations and then used for image reconstruction. When using EIT, an accurate detection of the aorta is important for calculation of hemodynamic parameters. A novel algorithm for the detection of aortic pixels in EIT images is presented. The accuracy of the algorithm was validated on animal data with intermodal (computed tomography, F1-score 0.25) and intramodal (injection of contrast solution, F-1 score 0.8125) ground truth. In a next step, EIT signal was measured on 5 healthy volunteers and the location of the aorta was successfully detected with the algorithm. By exploiting ECG triggered ensemble averaging a representative aortic BP waveform was acquired which allowed for detection of BP pulse arrival time (PAT). A strong negative linear correlation was observed between the mean arterial BP and PAT detected with EIT (r=-0.8593). The results of this work are quite promising and suggest that EIT assessment of hemodynamic parameters might be a viable alternative for current methods. The EIT technology is however still in its development phase and the various reconstruction algorithms available with their reconstruction settings have to be further tested on a larger scale followed by methodological in-vivo trials in order to pave its way to clinical use. A conference papers based on the results of this thesis was accepted for presentation at the 19th Electrical Impedance Tomography conference.
en
dc.language
English
-
dc.language.iso
en
-
dc.rights.uri
http://rightsstatements.org/vocab/InC/1.0/
-
dc.subject
Bioimpedanz
de
dc.subject
Biomedizinische Signalverarbeitung
de
dc.subject
Medizinische Signalerkennung
de
dc.subject
Bildrekonstruktion
de
dc.subject
Bioimpedance
en
dc.subject
Biomedical signal processing
en
dc.subject
Medical signal detection
en
dc.subject
Image reconstruction
en
dc.title
Detection of aortic hemodynamics in Electrical Impedance Tomography using functional regions of interest : investigation of in-vivo, in-silico and phantom measurements
en
dc.title.alternative
Detektion der Aorta-Hämodynamik in der elektrischen Feldplethysmographie mittels regionaler Analyse : in-vivo, in-silico und phantom Messungen
de
dc.type
Thesis
en
dc.type
Hochschulschrift
de
dc.rights.license
In Copyright
en
dc.rights.license
Urheberrechtsschutz
de
dc.identifier.doi
10.34726/hss.2018.53923
-
dc.contributor.affiliation
TU Wien, Österreich
-
dc.rights.holder
Martin Elenkov
-
dc.publisher.place
Wien
-
tuw.version
vor
-
tuw.thesisinformation
Technische Universität Wien
-
dc.contributor.assistant
Thürk, Florian
-
tuw.publication.orgunit
E354 - Electrodynamics, Microwave and Circuit Engineering
-
dc.type.qualificationlevel
Diploma
-
dc.identifier.libraryid
AC15026648
-
dc.description.numberOfPages
90
-
dc.identifier.urn
urn:nbn:at:at-ubtuw:1-110922
-
dc.thesistype
Diplomarbeit
de
dc.thesistype
Diploma Thesis
en
dc.rights.identifier
In Copyright
en
dc.rights.identifier
Urheberrechtsschutz
de
tuw.advisor.staffStatus
staff
-
tuw.assistant.staffStatus
staff
-
tuw.advisor.orcid
0000-0002-1228-3859
-
item.fulltext
with Fulltext
-
item.mimetype
application/pdf
-
item.openairecristype
http://purl.org/coar/resource_type/c_bdcc
-
item.cerifentitytype
Publications
-
item.openairetype
master thesis
-
item.grantfulltext
open
-
item.openaccessfulltext
Open Access
-
item.languageiso639-1
en
-
crisitem.author.dept
E307-03 - Forschungsbereich Biomechanik und Rehabilitationstechnik
-
crisitem.author.parentorg
E307 - Institut für Konstruktionswissenschaften und Produktentwicklung