<div class="csl-bib-body">
<div class="csl-entry">Resconi, A. (2019). <i>Instrumentation of an in-vitro test bed for load testing of shoulder endoprosthesis prototypes</i> [Diploma Thesis, Technische Universität Wien]. reposiTUm. http://hdl.handle.net/20.500.12708/79397</div>
</div>
-
dc.identifier.uri
http://hdl.handle.net/20.500.12708/79397
-
dc.description.abstract
Shoulder instability remains one of the major complications after total shoulder arthroplasty and reverse TSA. Prototypes of shoulder endoprostheses designed to tackle this problem were developed at the Ludwig Boltzmann Institute for Experimental and Clinical Traumatology. In this thesis the design and function of the shoulder simulator that was commissioned to test the prostheses are described. The simulator is an automated biaxial test bed that measures separation forces between the humerosocket and glenosphere in every physiological position of the glenohumeral joint, with a maximum load of 500 N. The mechanical apparatus is able to simulate two different decoupling mechanisms, along the sagittal plane and along the transverse plane, at different angles between the prosthesis partners. This thesis covers all steps of calculation, design and construction, but not the implementation of the software. The device is validated according to design, installation and operational qualification.
en
dc.format
vii, 124 Seiten, 29 ungezählte Seiten
-
dc.language
English
-
dc.language.iso
en
-
dc.subject
test bed
en
dc.subject
shoulder endopirostheses
en
dc.title
Instrumentation of an in-vitro test bed for load testing of shoulder endoprosthesis prototypes