Title: Vorklinische Verifikation eines Bestrahlungsplanungssystems mit Monte-Carlo Dosisberechnung und biologischer Optimierung
Language: English
Authors: Fotina, Irina 
Qualification level: Diploma
Advisor: Georg, Dietmar
Issue Date: 2008
Number of Pages: 80
Qualification level: Diploma
Abstract: 
The aim of this thesis was to investigate IMRT treatment planning with treatment planning systtem Monaco and to perform measurements for its commissioning and subsequent verification of the IMRT treatment plans. Monaco combines Monte Carlo dose calculation algorithm and biologically expressed objective functions. These two innovating features imply some specialties on commissioning measurements and planning process. Monte Carlo dose calculations aim to provide excellent level of accuracy for the dose calculation with insignificant prolongation of the planning time. Use of the biologically-based objective functions makes the planning process easy and transparent for any corrections and an EUD formalism of optimization constraints helps for the plan analysis and evaluation.
For the patient-specific plan validation we used dedicated phantoms, where IMRT plans were recalculated. The measurements with EBT films and ion chambers were performed and compared with the results from the treatment planning system. All results obtained with Monaco were inside the established acceptance criteria, which was also confirmed by verification on the independent software for monitor unit verification.
The acceptable dose deviation for ionization chamber measurements and MU verification should be below 3% and criteria for [gamma]-index verification also deal with 3 mm distance-to-agreement and 3% dose difference, expressed in terms of [gamma]mean and [gamma]>1.
The issues on planning and verification procedures together with respective results described in this thesis enable a safe implementation of such TPS into the clinical routine, but some follow-up dosimetric tests and further exploration of the system possibilities for different planning solutions are highly recommended.
Keywords: Bestrahlungsplanung; IMRT; Monte-Carlo Dosisberechnung; Biologische Optimierung; Verifikationsmessungen
treatment planning; IMRT; Monte-Carlo dose calculation; Biological optimization; Verification measurements
URI: https://resolver.obvsg.at/urn:nbn:at:at-ubtuw:1-25802
http://hdl.handle.net/20.500.12708/11021
Library ID: AC05037175
Organisation: E141 - Atominstitut 
Publication Type: Thesis
Hochschulschrift
Appears in Collections:Thesis

Files in this item:


Page view(s)

25
checked on Oct 9, 2021

Download(s)

74
checked on Oct 9, 2021

Google ScholarTM

Check


Items in reposiTUm are protected by copyright, with all rights reserved, unless otherwise indicated.