Ismail, S. S. (2010). A New Automated Sample Transfer System for Instrumental Neutron Activation Analysis. Journal of Automated Methods and Management in Chemistry, 2010, 1–8. https://doi.org/10.1155/2010/389374
Computer Science Applications; Instrumentation; Analytical Chemistry; Chemical Engineering (miscellaneous)
-
Abstract:
A fully automated and fast pneumatic transport system for short-time activation analysis was recently developed. It is suitable for small nuclear research reactors or laboratories that are using neutron generators and other neutron sources. It is equipped with a programmable logic controller, software package, and 12 devices to facilitate optimal analytical procedures. 550 ms were only necessary to transfer the irradiated capsule (diameter: 15 mm, length: 50 mm, weight: 4 gram) to the counting chamber at a distance of 20 meters using pressurized air (4 bars) as a transport gas.